Persistent Identifier
|
perma:LIST.PWKAOV |
|
Publication Date
|
2026-07-06 |
|
Title
| The Peregrine Ion Trap Mass Spectrometer (PITMS): Results from a CLPS-delivered Mass Spectrometer [* Cross-Reference *] |
|
Other Identifier
| https://doi.org/10.3847/PSJ/ad9927
SCOPUS_ID:85215837258 |
|
Author
| Cohen, Barbara A. (NASA Goddard Space Flight Center) - ORCID: 0000-0001-5896-5903
Barber, Simeon J. (The Open University)
Gawronska, Aleksandra J. (NASA Goddard Space Flight Center, College of Engineering, Physics, and Computing)
Abernethy, Feargus A.J. (The Open University)
Curran, Natalie M. (NASA Goddard Space Flight Center, College of Engineering, Physics, and Computing)
Driggers, Phillip A. (NASA Goddard Space Flight Center)
Farrell, William M. (Space Science Institute, DeepSpace Technologies Inc.)
Heather, David J. (ESTEC - European Space Research and Technology Centre) - ORCID: 0000-0001-7338-0225
Howe, Christopher (RAL Space)
Landsberg, Peter F. (The Open University)
López-Días, Veneranda (Luxembourg Institute of Science and Technology)
Morse, Andrew D. (The Open University) - ORCID: 0000-0001-9683-316X
Morse, Thomas (RAL Space)
Poston, Michael J. (Southwest Research Institute) - ORCID: 0000-0001-5113-1017
Prem, Parvathy (Johns Hopkins University Applied Physics Laboratory) - ORCID: 0000-0001-8369-8587
Trautner, Roland (ESTEC - European Space Research and Technology Centre) - ORCID: 0000-0002-5190-2894
Tucker, Orenthal J. (NASA Goddard Space Flight Center)
Warren, Tristram J. (University of Oxford)
Boccelli, Stefano (NASA Goddard Space Flight Center, Oak Ridge Associated Universities) |
|
Point of Contact
|
Use email button above to contact.
LIST RDS (LIST) |
|
Description
| The Peregrine Ion Trap Mass Spectrometer (PITMS) was a mass spectrometer designed to measure lunar gases. PITMS flew on the first flight of Astrobotic’s Peregrine lander via the Commercial Lunar Payload Services (CLPS) program in 2024 January. After launch, the lander suffered a propulsion system anomaly that prevented the mission from reaching the Moon, but PITMS collected 80 high-quality spectra while in cislunar space. PITMS observed abundant outgassing products from the Peregrine lander, including water, MON-25 oxidizer from the propulsion system leak, and traces of combustion products. PITMS data help constrain the nature of the propulsion system failure: oxidizer molecular ratios show that the leak released molecules rapidly enough for them to fully dissociate, and the high observed abundances imply that the oxidizer traveled within the lander surfaces rather than jetting into space. The amount of water offgassed by the spacecraft is substantially more than other planetary spacecraft, so the PITMS results suggest that instruments flying in the CLPS paradigm need to consider lander cleanliness. Though not successful in measuring the native lunar exosphere, the PITMS results showcase the capabilities of a mass spectrometer on board a lunar lander, along with lessons in pragmatism and flexibility that would enable such an instrument to ultimately be successful in the CLPS initiative. (2025-01-01)
***This entry has been automatically imported via Infodoc(ASO) CSV by LIST harvest scripts. Please refer to https://doi.org/10.3847/PSJ/ad9927 for the original and latest version of the dataset and data downloads*** (2026-06-09) |
|
Subject
| Earth and Environmental Sciences |
|
Keyword
| Peregrine Ion Trap
Ion Trap Mass
Ion Trap
Trap Mass Spectrometer
measure lunar gases |
|
Funding Information
| European Space Agency: 80GSFC21M0002 |
|
Deposit Date
| 2025-01-01 |
|
Data Type
| Article |
|
Data Source